FISHER & PAYKEL HEALTHCARE LTD INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT266
Device Problems Leak/Splash (1354); Cut In Material (2454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/25/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint rt266 infant dual heated evaqua2 breathing circuit was returned to fisher & paykel healthcare in (b)(4) for investigation.The complaint device was visually inspected.Results: visual inspection of the complaint circuit revealed cut-like damage to the inspiratory limb.A lot check revealed no other complaints of this nature for the lot number provided.Conclusion: the damaged inspiratory limb of the complaint circuit appears to have been damaged with a sharp object.All rt266 infant dual heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.This suggests that the damage occurred to the circuit after it was released for distribution.Our user instructions that accompany the rt266 infant dual heated evaqua2 breathing circuit state the following: check all connections are tight before use; perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient; set appropriate ventilator alarms; do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers.
 
Event Description
A hospital in (b)(6) reported via a distributor that the inspiratory limb of an rt266 infant dual heated evaqua2 breathing circuit was leaking after 4 or 5 days of use.No patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
173 technology dr. suite 100
irvine, CA 92618
8007923912
MDR Report Key5610392
MDR Text Key43716723
Report Number9611451-2016-00173
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 03/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT266
Device Catalogue NumberRT266
Device Lot Number150718
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2016
Date Manufacturer Received03/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-