SYNTHES RARON 3.5MM LCP® PROXIMAL HUMERUS PLATE-STANDARD 3H SHAFT/90MM; APPLIANCE, FIXATION, NAIL/BLADE/ PLATE COMBO, SINGLE COMPONENT
|
Back to Search Results |
|
Catalog Number 241.901 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Pain (1994); Fracture, Arm (2351)
|
Event Type
Injury
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.The subject device is not expected to be returned to the synthes manufacturer for evaluation and was reportedly disposed of by the hospital.A device history record review was performed for the subject device lot.The review showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.No non-conformances were generated during the production of the subject device.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the patient underwent revision surgery on (b)(6) 2016 to treat a humeral fracture located distally to a previously implanted 3.5mm locking compression plate (lcp) 90mm, three cortex screws and nine locking screws.It is unknown when the original hardware was implanted.It was reported that the patient was traveling on an unknown date and experienced pain when reaching.It is not known when the fracture occurred.During the (b)(6) 2016 revision surgery, the existing hardware was noted to be intact.The 3.5mm lcp humerus plate 90mm and two cortex screws were removed.The patient was revised to a longer 3.5mm lcp 196mm and the one of the original cortex screws and the nine original locking screws were re-implanted.The surgery was successfully completed without delay.This report is 1 of 3 for com-(b)(4).
|
|
Search Alerts/Recalls
|
|
|