| Catalog Number |
03756230 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/06/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the doctor found the screw slip during the lumbar spine surgery and completed the surgery changing to another one.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection; device history review; complaint history review; risk assessment; results: it was reported that the when the doctor tried to screw, the device slipped.No relevant manufacturing issues found.Poor alignment between blocker and hex driver may prevent blocker is to properly engage and may fall into wound.Conclusion: the possible root cause of this event is likely the poor alignment between the blocker and the hex driver and excessive force applied to the blocker.
|
| |
|
Event or Problem Description
|
|
It was reported that the doctor found the screw slip during the lumbar spine surgery and completed the surgery changing to another one.
|
| |
|