• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT KYPHON NEUCHATEL MFG KYPHON XPANDER¿ INFLATABLE BONE TAMP; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT KYPHON NEUCHATEL MFG KYPHON XPANDER¿ INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number K09A
Device Problem Material Rupture (1546)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 03/30/2016
Event Type  Injury  
Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Pre-op diagnosis: multiple myeloma it was reported that on (b)(6) 2016, intra-op during balloon kyphoplasty, balloon burst in the vertebral body after inflating, cement was filled up as it was, surgeon waited for the cement to harden and surgery was finished.No patient complications were reported.Balloon kyphoplasty was performed at three levels, th11- l1.The balloon got damaged when it was inflated around 300 psi at right th12.Then cement was inserted without cleaning the surgical field.It was unknown which lot was damaged.The patient had bone sclerosis and surgeon told bones were hard.No complication was reported, the fragment remained in the patient though.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER¿ INFLATABLE BONE TAMP
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
MDT KYPHON NEUCHATEL MFG
pierre-à-bot 97
neuchâtel,ne 2000
CH  2000
Manufacturer (Section G)
MDT KYPHON NEUCHATEL MFG
pierre-à-bot 97
neuchâtel,ne 2000
CH   2000
Manufacturer Contact
greg anglin
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key5614317
MDR Text Key43849526
Report Number2953769-2016-00032
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K041454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberK09A
Device Lot Number193058-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-