An analysis was performed on the returned lead portions and the reported allegations of "fracture of leads" were confirmed.Note that the (-) green electrode was not returned for analysis and therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis of the returned 40mm portion the (+) white electrode quadfilar coil appeared to be broken approximately 19mm from the electrode bifurcation.Scanning electron microscopy was performed and identified the area as having extensive pitting which prevented identification of the coil fracture type.Pitting was observed on the coil surface.During the visual analysis of the returned 9mm portion the end of the (+) white electrode quadfilar coil appeared to be broken approximately 18mm past the end of the cut inner silicone tubing.Scanning electron microscopy was performed and identified the area on two of the broken coil strands as having evidence of a stress induced fracture (fatigue appearance) with mechanical damage and pitting.The area on the remaining broken coil strands was identified as being mechanically damaged which prevented identification of the coil fracture with pitting.Flat spots and pitting were observed on the coil surface.It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.Low magnification sem analysis of the quadfilar coil shows characteristics typical of a lead discontinuity which may include: material fracture, rough or pitted surface, thinned material thickness, electro-etching or material dissolution.What appeared to be white deposits were observed in various locations.Eds (energy dispersion spectroscopy ¿ provides chemical or element identity/composition analysis) was performed on the deposit observed on the outer silicone tubing and identified the deposit as containing silicon, phosphorus, sodium, magnesium and calcium.During the visual analysis the (+) white electrode ribbon appeared to be embedded in what appeared to be remnants of dried body tissue.This condition may have prevented the (+) white electrode ribbon from coming in contact with the vagus nerve.With the exception of the observed discontinuity and the tissue-covered (+) white electrode ribbon the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pins provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no other discontinuities were identified.Based on the findings in the product analysis lab, there is evidence to suggest a discontinuity in the returned portions of the device which may have contributed to the stated allegations of high impedance.Note that since the (-) green electrode was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.However, the positive electrode condition may have contributed to the reported ¿high impedance¿.
|