| Model Number |
BA25-80/I20-40 |
| Medical Device Problem Codes |
Failure To Adhere Or Bond (1031); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/30/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device involved in the event will not be returned for evaluation as it remains implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device remains implanted in the patient.
|
| |
|
Event or Problem Description
|
|
Patient initially implanted with a bifurcated stent and an infrarenal aortic extension on (b)(6) 2016.Upon follow up, computed tomography (ct) showed a partial stent collapse and a type 1a endoleak.The physician elected to use a stiff wire to alter the initial infrarenal device and implanted a suprarenal aortic extension to seal the endoleak.The patient is in stable condition.
|
| |
|
Search Alerts/Recalls
|