• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALPHATEC SPINE INC PEDICLE PROBE, LENKE CURVED (STAINLESS STEEL); HXB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALPHATEC SPINE INC PEDICLE PROBE, LENKE CURVED (STAINLESS STEEL); HXB Back to Search Results
Model Number 22SPT10
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/01/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information provided by the sales rep revealed the patient had really hard dense bone.A mallet was used to advance the probe into the patients pedicle when it broke.The detached tip of the probe was removed from the patient without issue.An evaluation of the returned lenke curved probe found the instrument had broken approximately 35mm from the distal tip.The 35mm location is the transition point where the 6.35mm diameter shaft tapers down to the pointed curved tip.The distal pointed curved tip contains the smallest diameter on the instrument, and is subject to the most stress during use.Bone probes are designed to locate the proper pathway and length of the screw which is to be implanted.When used as intended the instrument would not come in contact with the amount of force required to fracture and/or deform the tip in this manner.
 
Event or Problem Description
The tip of curved probe fractured and broke during use.The event caused no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEDICLE PROBE, LENKE CURVED (STAINLESS STEEL)
Common Device Name
HXB
Manufacturer (Section D)
ALPHATEC SPINE INC
5818 el camino real
carlsbad CA 92008
MDR Report Key5624718
Report Number2027467-2016-00025
Device Sequence Number16147297
Product Code HXB
UDI-Device Identifier00840967190112
UDI-Public00840967190112
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number22SPT10
Device Catalogue Number22SPT10
Device Lot Number6505607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2016
Initial Report FDA Received Date05/03/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-