Catalog Number 317-09
Device Problem Output below Specifications (3004)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, a device history record could not be reviewed.The instructions for use states the following: "6.Once the pads are primed, assure the flow rate displayed on the control panel is greater than 2.3 liters per minute, which is the minimum flow rate for a fullpad kit.¿ (b)(4).Sample was discarded.
 
Event Description
It was reported that the pads were giving a low flow.Therapy was interrupted in order to replace the pads with a new set of pads.Therapy was resumed with a new set of pads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTICGEL PADS
Type of Device
ARCTIC SUN GEL PADS
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
janna parks
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key5627690
MDR Text Key44778615
Report Number1018233-2016-00566
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002577
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number317-09
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/06/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-