SYNAPSE BIOMEDICAL INC. NEURX DIAPHRAGM PACING SYSTEM; DIAPHRAGMATIC/PHRENIC NERVE LAPAROSCOPICALLY-IMPLANTED STIMULATOR Back to Search Results
Model Number 20-0035
Medical Device Problem Code No Pacing (3268)
Health Effect - Clinical Codes Death (1802); Hypoxia (1918); Brain Injury (2219)
Type of Reportable Event Death
Additional Manufacturer Narrative
The device associated with this event was sent directly from the user facility to a certified, independent laboratory ((b)(6)) for testing.The testing was performed in accordance with a protocol provided by synapse biomedical inc.The protocol and summary of the results obtained are provided.The final report has not been completed but will be submitted in a follow-up report, which will contain details for each test conducted.Synapse biomedical inc.Has no further details regarding the adverse event other than those presented in the user facility report.We have had no contact with the patient's family or anyone else directly involved and it should be noted that the patient was not at the user facility at the time of the incident.The only mechanism for device shutdown is a mechanical switch that requires two buttons on the front of the device to be pressed simultaneously.The returned device was found to operate within specifications and the events described in the user facility report could not be reproduced.Manufacture and test records indicate the device was assembled according to controlled, predefined procedures and all testing met specification.There were no errors or deviations found in the device history record nor any indication that the device was defective in any way.A final conclusion will be included with the follow-up report but at this time we have identified no defects with the device, it performs within specifications and we have not be able to reproduce the adverse event as described in the user facility report.
 
Event or Problem Description
The patient was implanted with the synapse biomedical diaphragm pacing system (dps) device (b)(6) 2015.The implanting surgeon contacted a synapse biomedical representative on (b)(6) 2016 reporting that the patient had died.He stated that the mother of the patient had left the patient alone while using the dps and off the mechanical ventilator.When she came back the patient was not breathing and the dps was off.Additional information can be found in user facility report number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURX DIAPHRAGM PACING SYSTEM
Common Device Name
DIAPHRAGMATIC/PHRENIC NERVE LAPAROSCOPICALLY-IMPLANTED STIMULATOR
Manufacturer (Section D)
SYNAPSE BIOMEDICAL INC.
300 artino street
oberlin OH 44074
MDR Report Key5628922
Report Number3005868392-2016-00001
Device Sequence Number18968458
Product Code OIR
UDI-Device Identifier00852184003281
UDI-Public(01)00852184003281
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
H070003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type User Facility
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/30/2016
Device Model Number20-0035
Device Lot Number20-0035-102815-3-2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/03/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age20 YR
-
-