Model Number 403713
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/29/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event has been determined to be a post-placement/pre-load implant failure.This implant loss suggests that the source of the problem was likely to stem from procedural errors and/or the lack of osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, or patient behavior.Proper intended use of the implant is described in its instructions for use.Physical analysis was not performed.
 
Event or Problem Description
Implant failure due to mobility and infection.No osseointegration 3-6 months before discovery.Patient has cancer.Bone loss was also reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTERNAL HEX IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ACE SURGICAL SUPPLY COMPANY
1034 pearl street
brockton MA 02301
MDR Report Key5641290
Report Number1287163-2016-50433
Device Sequence Number49164
Product Code DZE
UDI-Device Identifier00614950005437
UDI-Public00614950005437
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K954513
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2016
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Expiration Date03/30/2019
Device Model Number403713
Device Catalogue Number403713
Device Lot Number14040020
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2016
Initial Report FDA Received Date05/09/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-