MERGE HEALTHCARE MERGE CARDIO; PICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS Back to Search Results
Model Number MERGE CARDIO V9.0.8
Device Problem Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/06/2016
Event Type  malfunction  
Manufacturer Narrative
The issue was being manually corrected by users at the facility.Further investigation of the issue determined that the measurement was incorrectly mapped at the time of implementation.Merge healthcare support corrected the mapping, and confirmed with the facility that the correct value is now imported.
 
Event Description
Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.On (b)(6) 2016, a customer contacted merge healthcare and stated that a diagnostic measurement taken by their modality was importing into the incorrect node within the clinical reporting tool.Due to an incorrect value displaying in the diagnostic report, there is a potential for incorrect treatment of a patient that could result in harm.There is no indication that any actual patient harm.Reference complaint number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE CARDIO
Type of Device
PICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge dr.
hartland, WI 53029
2629123570
MDR Report Key5643015
MDR Text Key44893733
Report Number2183926-2016-00580
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Health Professional
Remedial Action Modification/Adjustment
Type of Report Initial
Report Date 04/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE CARDIO V9.0.8
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/12/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-