• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIODYNAMICS; STOPCOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIODYNAMICS; STOPCOCK Back to Search Results
Catalog Number H965700550031
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 03/02/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lot met all material, assembly and performance specifications.The angiodynamics (b)(4) 2016 complaint report was reviewed for the stopcocks product family and the failure mode,"hole in pouch." no adverse trends were identified.The returned sample was visually inspected and the hole in the pouch was confirmed.There was a nipro label on the pouch with the hole.The root cause of the hole is the handle from the stopcock presses outward against the tyvek.As part of the receiving process at nipro (the distributor in (b)(4)), all pouched products are removed from their inner boxes and a nipro label (in (b)(4)) is applied to each pouch.Additional handling may occur if product is 100% visually inspected.The pouched product is then re-boxed into the inner box by the nipro warehouse employees.The hole in the tyvek was likely caused by the handle of a stopcock in the pouch, however, what caused the handle to be pushed against the tyvek in a manner that resulted in a hole cannot be determined.Potential contributing factors include: handling of the pouches as they are placed in the inner boxes.Handling during transit to nipro warehouse.Handling during nipro labeling/inspection and re-boxing process.All pouches are 100% inspected per angiodynamics procedures during the sealing and final box processes.The directions for use (dfu) packaged with the kits contain the following warning: "contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged." (b)(4).The mdr for this complaint record was originally submitted on (b)(4) 2016 via usps.Due to e-submitting requirements, it was returned for resubmitting.
 
Event or Problem Description
As reported by angiodynamics' distributor in (b)(6), in the distributor's warehouse a small hole was found in the tyvek portion of a stopcock kit pouch, breaching the sterility.The device had not been provided to a hospital and was returned to angiodynamics for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS
Common Device Name
STOPCOCK
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key5644390
Report Number1317056-2016-00041
Device Sequence Number337012
Product Code DTL
UDI-Device IdentifierH965700550031
UDI-PublicH965700550031
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K842829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/31/2018
Device Catalogue NumberH965700550031
Device Lot Number4963506
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2016
Initial Report FDA Received Date05/10/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-