Catalog Number 72200616
Device Problems Overheating of Device (1437); Self-Activation or Keying (1557); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/21/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the powermax elite hand control overheated and became very hot.The powermax elite hand control reportedly got very hot and was possibly running on it's own.A back up device was used to finish the case with no further issues.It was reported that there were no patient or staff injuries associated with the alleged event.A short delay of less than 30 minutes was reported.
 
Manufacturer Narrative
Functional evaluation confirmed the reported complaint of overheating.Unit's housing heats up and motor stalls giving a blade stall error during testing.The cable assembly was removed and unit was observed to be corroded internally.The root cause of the overheating is most likely internal corrosion of gearbox.A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records found no inconsistencies.There are no indications that would suggest the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key5645555
MDR Text Key44905598
Report Number1643264-2016-00068
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/22/2016
Initial Date FDA Received05/10/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/29/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-