| Catalog Number |
4823717580 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/18/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; tulip head broke off screw while inserting blocker.Update: all of the broken fragments were retrieved from the surgical site.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Unanticipated hazard.Method: device history review; complaint history review; risk assessment; results: manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.The reported event occurred while doctor was using a standard blocker inserter.No persuader was being used by the surgeon according to communication with the sales representative.Surgical technique states the blocker is assembled onto the universal tightener for insertion.Conclusion: a definitive root cause of the event could not be determined with the given information.
|
| |
|
Event or Problem Description
|
|
It was reported that; tulip head broke off screw while inserting blocker.Update: all of the broken fragments were retrieved from the surgical site.There were no adverse consequences to the patient.
|
| |
|