• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XACT CAROTID STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XACT CAROTID STENT SYSTEM Back to Search Results
Catalog Number 82087-01
Device Problem Inaccurate Delivery (2339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/18/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).Evaluation summary: visual inspection was performed on the returned device.The inaccurate delivery was not confirmed as the stent had already been deployed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation was unable to determine a conclusive cause for the reported inaccurate delivery.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event Description
It was reported that during the external carotid procedure with mild calcification, in the vessel the 9.0 x 30 mm xact stent was being deployed when the stent jumped approximately 15 mm and covered only a distal portion of the target lesion.A second xact stent was used as treatment of the lesion.There was no reported adverse patient sequela.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XACT CAROTID STENT SYSTEM
Type of Device
CAROTID STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key5648968
MDR Text Key45052415
Report Number2024168-2016-03072
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 05/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2018
Device Catalogue Number82087-01
Device Lot Number5110461
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2016
Initial Date FDA Received05/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
-
-