| Model Number |
MICL13.2 |
| Medical Device Problem Codes |
Difficult To Position (1467); Inadequacy of Device Shape and/or Size (1583); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Code Available (3191)
|
| Date of Event |
04/07/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter stated the surgeon inserted a 13.2mm micl13.2 implantable collamer lens, -5.0 diopter, and noted the lens was too long for the patient's eye.The lens was removed within the same surgery with no patient injury.Another lens was not implanted.The surgeon felt the lens may have been mislabeled.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information received - the reporter indicated the lens was inserted and removed from the left eye (os).The incision was enlarged to remove the lens and sutures were required to close the incision.The reporter indicated the icl would not position correctly and kept rotating out of position.During removal, the icl caught the iris which caused depigmentation and atrophy temporally.The patient experienced photosensitivity.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: device history record review.Result: (no failure detected): based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within the established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Conclusion: (unable to confirm complaint): based on the complaint history, work order search, medical review, device history record review and product evaluation, a specific root cause of the event could not be determined.(b)(4).
|
| |
|