MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA CATHETER; PTA DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA CATHETER; PTA DILATATION CATHETER Back to Search Results
Catalog Number U3575625
Device Problems Break (1069); Leak/Splash (1354); Retraction Problem (1536)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 02/17/2016
Event Type  malfunction  
Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the pta balloon catheter hub allegedly leaked during the first inflation in the right sfa.There was no reported retraction difficulty through the sheath.It was further reported that another balloon was used to complete the procedure.There was no reported patient injury.
 
Manufacturer Narrative
Manufacturing review: a manufacturing review was conducted.The lot met all release criteria.Visual inspection: the sample was returned.The balloon size for this product is printed on the balloon hub of the catheter and identified the returned sample as a 6mm x 250mm balloon.The balloon did not appeared to have been previously inflated.No anomalies were noted to the strain relief or the y-hub.Functional/performance evaluation: the patency was tested using an in-house.035¿ guidewire and passed without issue.The inflation hub was connected to an inflation device and an attempt was made to inflate the balloon with water.Upon inflation, water was observed to be leaking out the distal end of the strain relief.The strain relief was removed and a partial circumferential catheter shaft break was observed at the distal end of the y-hub.Sanding marks were noted on the catheter underneath the strain relief and at the location of the break.Medical records review: as medical records were not provided, a review could not be performed.Image/photo review: no medical images have been made available to the manufacturer.Conclusion: the investigation is confirmed for a partial circumferential catheter shaft break just distal to the y-hub, resulting in the reported leak.Excessive sanding of the catheter under the strain relief is the root cause for the partial break in the catheter.Labeling review: the current ifu (instructions for use) states: precautions: carefully inspect the catheter prior to use to verify that catheter has not been damaged during shipment and that its size, shape, and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.Use of the ultraverse 035 pta dilatation catheter: position the balloon relative to the lesion to be dilated, ensure the guidewire is in place, and inflate the balloon to the appropriate pressure.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the pta balloon catheter hub allegedly leaked during the first inflation in the right sfa.There was no reported retraction difficulty through the sheath.The procedure was concluded with the same device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA CATHETER
Type of Device
PTA DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI  
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key5655187
MDR Text Key46474179
Report Number2020394-2016-00442
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142261
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2016
Device Catalogue NumberU3575625
Device Lot Number50119471
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/19/2016
Initial Date FDA Received05/13/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
Patient Weight75
-
-