• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST Back to Search Results
Model Number 100071
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Issue (2379)
Patient Problems Transient Ischemic Attack (2109); Blurred Vision (2137); No Consequences Or Impact To Patient (2199)
Event Date 04/17/2016
Event Type  Injury  
Manufacturer Narrative
Investigation conclusion: investigation pending.
 
Event Description
Report received of discrepant inratio value.(b)(6).Patient went to hospital due to blurred/double vision on (b)(6) 2016.Patient admitted to hospital for tia.Released from hospital on (b)(6) 2016, no medication given, no changes in medication.Patient's therapeutic range 2.0-3.0 prior to hospitalization and tia.Patient's new therapeutic range 2.5-3.5.
 
Manufacturer Narrative
Investigation/conclusion: it is indicated that the product is not returning for evaluation.Therefore, a review of the entire in-house testing history of the lot was performed.In-house strip testing on the reported strip lot meets release criteria.The customer's complaint was not replicated and no product deficiency was identified.A review of the manufacturing batch records for the reported strip lot did not uncover any non-conformances.The lot met release specifications.Root cause could not be determined from the information provided by the customer.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INRATIO PT/INR TEST STRIPS
Type of Device
PROTHROMBIN TIME TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge road
san diego CA 92121
Manufacturer Contact
ya-ling king
9975 summers ridge road
san diego, CA 92121
8588052084
MDR Report Key5660128
MDR Text Key45334171
Report Number2027969-2016-00320
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number100071
Device Lot Number384368A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNSPECIFIED MEDICATIONS
Patient Outcome(s) Hospitalization;
-
-