• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC BANDER URETERAL DIVERSION STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC BANDER URETERAL DIVERSION STENT SET; FAD STENT, URETERAL Back to Search Results
Catalog Number 025227-OE
Medical Device Problem Code Occlusion Within Device (1423)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/07/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pt identifier) unknown as not provided.Age at time of event, date of birth) unknown as not provided.Weight) unknown as not provided.(b)(4).The event is currently under investigation.
 
Event or Problem Description
The female patient had the ureteral stent placed on (b)(6) 2016.The procedure went well.After two days they realized that the stent had been engorged by litho-problems.The user facility removed the stent and had to re-operate on the patient.When the doctor re-opened the patient, he decided to use another manufacturer's catheter.A section of the device did not remain inside the patient's body.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.The patient is reported to be doing well.
 
Additional Manufacturer Narrative
Investigation - evaluation a review of the complaint history, device history record, quality control, and specification of the device was conducted during the investigation.Device was not returned for investigation.There is no indication that a design or process related failure mode contributed to this event.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.A review of device history record did not observe any nonconformance that may have contributed to this incident.Based on the provided information a definitive root cause cannot be established or reported at this time.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.Per the risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BANDER URETERAL DIVERSION STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key5660892
Report Number1820334-2016-00316
Device Sequence Number13457945
Product Code FAD
UDI-Device Identifier00827002180703
UDI-Public(01)00827002180703(17)180830(10)6165223
Combination Product (Y/N)N
Initial Reporter CountryHU
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number025227-OE
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/07/2016
Device Age8 MO
Event Location Hospital
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/25/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/17/2016
Supplement Report FDA Received Date05/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-