| Model Number |
24-025-50-09 |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
04/25/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This is one of two related mdrs.Please refer to mdr 9610905-2016-00016.
|
| |
|
Event or Problem Description
|
|
A sternal plate broke.The patient underwent a secondary surgery to replace the plate.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was optically assessed and stereo microscopically investigated in the lab.The stereo microscopic investigation revealed a tensile crack due to mechanical overload.Further observation determined there were no indications of material or manufacturing defects discovered.The product history device records were also reviewed based on the lot number provided and revealed no abnormalities.The results of the investigation conclude that the root cause for breakage was due to mechanical overload on the device.If further information can be gathered that might add value to the contents of the investigated report, an additional follow-up report will be submitted.
|
| |
|