Model Number 701044054UL
Device Problems Melted (1385); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A maquet field service technician was on site and investigated the unit.The technician troubleshot the device and found corroded valves on the main side of the device.The shunt valve and 3-way valve were replaced and the unit was tested for functionality.A test run was initiated.All tests were performed successfully.A supplemental medwatch will be submitted if additional information becomes available.
 
Event Description
It was reported that the complete ice block of a hcu40 melted down during surgery.The incident occurred during patient treatment.No negative consequences were reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU40 DEVICE
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key5663452
MDR Text Key46362289
Report Number8010762-2016-00340
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K130300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701044054UL
Device Catalogue Number701044054UL
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/26/2016
Initial Date FDA Received05/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-