ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEM08080
Medical Device Problem Codes Difficult or Delayed Positioning (1157); Difficult to Remove (1528)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/22/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.
 
Event or Problem Description
It was reported that during placement of the endovascular stent graft in the cephalic vein via access through a left upper arm a/v graft for treatment of a reoccurring lesion with a 30 % residual stenosis after balloon dilation, the stent graft could not be deployed.Reportedly, the anatomy of the tracking vessel was tortuous and the proximal end of the stent graft was placed in a corner.During retraction of the delivery system, difficulties occurred and the shaft appeared to be stretching.Reportedly, the entire system was removed.No other stent was placed this day.There was no reported patient injury.
 
Additional Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.The condition of the returned device confirmed the reported deployment difficulties.The outer sheath was found to be elongated which indicates that increased friction must have affected on the delivery system during the attempt to release the stent graft.Reportedly, the stent graft could be successfully implanted after the initial deployment failure and was not returned.The reported difficulties in removing the delivery system could not be reproduced.Except for the elongation of the outer sheath, the delivery system was found to be in good condition and a patency test with a device compatible 0.035"' guide wire could be performed successfully.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.The event may be associated with difficult anatomic conditions leading to increased friction during deployment.Reportedly, the anatomy of the tracking vessel was tortuous and the proximal end of the stent graft was placed in a corner.Insufficient flushing of the device may be another contributing factor to the reported event.On the basis of the information available, a definite root cause for the reported event could not be determined.The ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device of the same size." and "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." furthermore, the ifu states: "do not kink the delivery catheter or use excessive force during delivery to the target lesion." the ifu indicates that the device must be flushed with sterile saline.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
MDR Report Key5665291
Report Number9681442-2016-00134
Device Sequence Number13497888
Product Code PFV
UDI-Device Identifier04049519008509
UDI-Public(01)04049519008509(17)180723(10)ANZG2625
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P130029/S002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/23/2018
Device Catalogue NumberFEM08080
Device Lot NumberANZG2625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/27/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/18/2016
Supplement Report FDA Received Date08/01/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient Weight64
-
-