MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® LIFEBAND; CHEST COMPRESSION ASSEMBLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® LIFEBAND; CHEST COMPRESSION ASSEMBLY Back to Search Results
Model Number 8700-0701
Device Problems Break (1069); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2016
Event Type  malfunction  
Manufacturer Narrative
The autopulse lifeband in complaint will not be returned for investigation.Therefore, a physical investigation will not be performed.A supplemental report will be filed if the product is returned and investigation has been completed.The autopulse user guide warnings state: "do not strap across, or otherwise constrain the lifeband.Constraining the movement of the lifeband can damage or break the lifeband." the instructions also state that "if the lifeband cannot be closed, use manual cpr.
 
Event Description
It was reported that during patient use the autopulse platform lifeband blew open during compressions.The user adjusted the lifeband and restarted compressions.Compressions were performed (amount unknown) until the lifeband blew open again and a "pop" sound was heard.The yellow clip that align the lifeband together (alignment tab) had broken off.The user had to revert to manual cpr.The customer confirmed that there was no malfunction of the autopulse.No patient sequelae was reported.No additional information was provided.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® LIFEBAND
Type of Device
CHEST COMPRESSION ASSEMBLY
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key5667951
MDR Text Key45575535
Report Number3010617000-2016-00343
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111001571
UDI-Public00849111001571
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8700-0701
Device Catalogue Number8700-0701
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-