The customer received questionable carbohydrate antigen 19-9 results for one patient when compared to other platforms.The data provided for the patient was: during a medical checkup, the result was "the upper normal range".In "bio medika", the result was 30.96 u/ml.In lab 1 with a cobas e601 analyzer, the result was 32 u/ml.The patient was checked again and the result on a siemens analyzer was 62 u/ml.In lab 2 with a cobas e601 analyzer, the result was 27 u/ml.In another laboratory with an unknown analyzer, the result was 53.5 u/ml.Information regarding if all of the results were from the same sample was requested, but was not provided.Information regarding if any erroneous results was reported outside the laboratory or if the patient was adversely affected was requested, but was not provided.The serial number (b)(4) was provided, but it was not clear which cobas e601 this was.The serial number of the other cobas e601 was requested, but was not provided.
|
A specific root cause could not be identified.Based on the provided, the results between roche analyzers were stable and comparable.Review of the provided calibration and qc data did not indicate any analyzer or reagent issue.Ca 19-9 results can vary when the same sample is tested with different assays from different manufacturer's and cannot be directly compared with one another.This difference is documented in product labeling.It is recommended to continue with one method for monitoring of the patient, or running several parallel measurements with the same sample on the different methods to confirm if the differences between the methods remain stable.
|