It was reported that a patient had high impedance on a follow-up visit with the physician that occurred on (b)(6) 2016.The patient also reported not feeling stimulation and an increase in seizures.In (b)(6), the patient had experienced a hard fall while hiking and the increase in seizures and the change in perception of the stimulation occurred after the fall.The seizure rate increased from 2-3 per month to 9 seizures during the month of april, after the fall.X-rays were sent to manufacturer for review.Based on the images provided, the source of the reported high impedance could not be determined.There did not appear to be any obvious lead discontinuities in the portion of the lead that could be visualized; however, part of the lead was difficult to assess due to the quality of the images.The presence of a micro-fracture in the lead could not be ruled out.Review of the x-rays showed inadequate strain relief.The patient underwent full revision surgery on (b)(6) 2016.Post-op system diagnostics showed impedance values within normal limits.No additional relevant information has been provided to date.
|
The generator and lead were received for analysis.Analysis of the lead was completed on 06/17/2016.Note that the electrodes were not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis what appeared to be pitting was observed on the connector pin surface.A definite cause for the pitting could not be determined based on the lead portion returned.The condition of the returned lead portion is consistent with that which typically exists following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portion were performed, during the visual analysis, and no discontinuities were identified.Analysis of the generator was completed on 06/21/2016.Review of the ram/flash data downloaded from the pulse generator shows a possible indication of increased impedance; the ¿diagvinitialprechange¿ value of 2085 ohms, the ¿diagvinitialpostchange¿ value of 19121 ohms, and the time of change detection (b)(6) 2016.It is likely that the generator was tested with a test resistor after explant to confirm high impedance within the lead.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.There were no additional performance or any other type of adverse condition found with the pulse generator.
|