Catalog Number 101-10001
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code No Information (3190)
Date of Event 06/16/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A comprehensive investigation was immediately initiated on receipt of the complaint.The product was not returned to manufacturer for evaluation, and a thorough investigation could not be completed as no lot number has been identified/confirmed in this case.Since the screw remains in the patient no physical, material, chemical evaluation could be performed, and the exact cause of the reported issue could not be ascertained.A review of the manufacturing and inspection records did not reveal any contributing information/trends.Without the implants, it is difficult to determine whether or not the rod was fully reduced or if the set screw may have been cross-threaded.The torque handles were returned and determined to be within specification.If more information becomes available manufacturer will file a follow-up report at that time.Remains in patient.
 
Event or Problem Description
It was reported to k2m, inc.On (b)(6) 2016 that two set screws backed out at the distal end of the construct post-operatively.
 
Additional Manufacturer Narrative
A comprehensive investigation was immediately initiated on receipt of the complaint.A review of all applicable material, inspection, manufacturing, storage and distribution records according to the description of the product used was conducted.All records revealed that the product lot was manufactured within specifications and distributed in accordance with all operating procedures.A review of the manufacturing and inspection records did not reveal any contributing information/trends.The set screw exhibited ideal "windshield wiper" type abrasions, which indicate an optimally seated and reduced rod, and therefore, ideal contact between the rod and set screw.While the threads exhibited signs of use, no major damage was observed.There were also no signs of cross-threading.The screw density was below one screw per level, which may have placed more stress on the set screw and contributed to it backing out.However, no root cause(s) could be determined.
 
Event or Problem Description
It was reported to k2m, inc.On (b)(6) 2016 that two set screws backed out at the distal end of the construct post-operatively.Patient was revised (b)(6) 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI SPINAL SYSTEM
Common Device Name
PEDICLE SCREW SPINAL SYSTEM
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
MDR Report Key5679127
Report Number3004774118-2016-00030
Device Sequence Number213213
Product Code NKB
UDI-Device Identifier10888857003842
UDI-Public10888857003842
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K042635
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number101-10001
Device Lot NumberDRJT AND DTWE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/29/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/25/2016
Supplement Report FDA Received Date07/20/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age15 YR
-
-