• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC. SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 2D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC. SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 2D Back to Search Results
Model Number SDM-00001-2D
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/20/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One of the footswitches that move the c-arm vertically had stuck.Both footswitches were replaced.
 
Event or Problem Description
C-arm lowered with patient in compression.The technologist was in the process of positioning the patient when the c-arm started to move down.The technologist moved the patient back out of the way.No injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 2D
Common Device Name
MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC.
36 & 37 apple ridge road
danbury CT 06810
MDR Report Key5680807
Report Number1220984-2016-00010
Device Sequence Number149077
Product Code OTE
UDI-Device Identifier15420045500013
UDI-Public15420045500013
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
PO80003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Remedial Action Repair
Type of Report Initial
Report Date (Section B) 05/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Medical Technologist
Device Model NumberSDM-00001-2D
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/25/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-