| Catalog Number |
31251180S |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/05/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported by the nurse of the hospital, that the locking pin 3125-1180s was missing.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation revealed the nail kit of the provided package to be the primary product (a different nail kit was originally complained - cat# 3125-1180s).No deviations were found during review of the manufacturing and inspection documents (dhr).The nail kit of the provided package labelling was documented as faultless prior to distribution; the set screw was documented as inserted.The microscopically inspection of the provided plastic blister revealed that a shape of a set screw is visible; furthermore micro-damaged, caused by the set screw thread windings, hitting the plastic during transport movement.Therefore a set screw was inserted within the blister; the complained missing of the set screw could not be confirmed.Most likely the set screw popped out of the package due to rough opening or due to opening upside down (use error).Review of complaint history, capa databases and risk analysis did not identify any discrepancies.There are no open actions in place related to the reported event for the subject product(s).No non-conformity was identified.
|
| |
|
Event or Problem Description
|
|
It was reported by the nurse of the hospital, that the locking pin 3125-1180s was missing.
|
| |
|