• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMPONENT, CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMPONENT, CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5515-F-602
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/06/2016
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
Surgical tech was unable to maintain sterility of implant in sterile packaging.
 
Manufacturer Narrative
An event regarding packaging issue involving a triathlon ps fem component, cemented was reported.The event was not confirmed.Method & results: -device evaluation: the packaging box with a shrink wrap partially open, package insert and the implant were returned.The tyvek lid of the inner and outer blister were not returned and so does the foam.However, there is evidence that there was a seal in the inner and outer blisters.The implant looks unremarkable.-clinician review: not performed as medical records were not received for review with a clinical consultant.-device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.-complaint history review: there have been no similar previous reported events for this lot id.Conclusions: the reported event could not be confirmed as the condition of the lids for both the inner and outer blisters were not returned.There is evidence of seal on both blisters.The packaging box looks intact and no signs of damage were observed.If additional information are received, this investigation will be reopened and re-evaluated.
 
Event Description
Surgical tech was unable to maintain sterility of implant in sterile packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS FEM COMPONENT, CEMENTED
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5693132
MDR Text Key47305373
Report Number0002249697-2016-01780
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number5515-F-602
Device Lot NumberUTUYD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received07/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-