| Model Number |
CC4204A |
| Medical Device Problem Code |
Split (2537)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/10/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Brand name: collamer® ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.Device evaluated by manufacturer: no, lens discarded.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter stated the surgeon inserted a cc4204a collamer single piece lens, +22.0 diopter, and the lens split in two after insertion into the eye.The lens was removed with no patient injury and the backup lens was implanted.The cause of the event was due to the lens was loaded incorrectly by the scrub technician.The facility discarded the lens.
|
| |
|
Additional Manufacturer Narrative
|
|
Method code(s): (process evaluation): device history record review.Result code(s): (no failure detected): based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within the established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.The cause of the event was due to the lens was loaded incorrectly by the scrub technician.(b)(4).
|
| |
|