Data obtained and decoded from the user's programming system confirmed that an end of service indicator had been observed following normal readings.An end-of-service warning message was verified during testing and was found to be associated with the output being disabled by the pulse generator.Burn marks were observed on the pulse generator case, which indicated that the pulse generator was likely exposed to an electro-cautery tool during device implant/explant.A reset of the pulsedisabled bit in the pulse generator memory was performed to allow for an output to once again be provided by the pulse generator for subsequent testing.The pulse generator diagnostics were as expected for the programmed parameters.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery, 3.066 volts as measured during testing shows an ifi=no condition.The data in the diagaccumconsumed memory locations revealed that 3.082% of the battery had been consumed.Other than the noted event (pulsedisabled), there were no additional performance or any other type of adverse conditions found with the pulse generator.
|