• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. CVC SET: 3-LUMEN 7 FR X 20 CM; ARROWG+ARD CATHETER PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. CVC SET: 3-LUMEN 7 FR X 20 CM; ARROWG+ARD CATHETER PRODUCTS Back to Search Results
Catalog Number CS-25703-E
Device Problems Kinked (1339); Physical Resistance (2578)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/01/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).No sample will be returned for evaluation.
 
Event Description
It was reported that in the icu, air bubbles were found in the arrow raulerson syringe when aspirating.As a result, new a kit was opened and used without issue.A delay was reported, however, there was no additional harm to the patient due to this delay or as a result of this occurrence.
 
Manufacturer Narrative
(b)(4).Device evaluation: the report air bubbles were found in the ars when aspirating could not be confirmed.Returned by the customer was one arrow raulerson syringe (ars).A visual inspection of the ars was performed and was typical in appearance.Microscopic inspection was performed, no defect in the syringe were noted.Functional testing was performed by aspirating water into the ars multiple times; water filled the syringe with no signs of leakage or air bubbles observed.A device history records review was performed and found no evidence to suggest a manufacturing related issue.No problem was found on the returned sample.No further action will be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CVC SET: 3-LUMEN 7 FR X 20 CM
Type of Device
ARROWG+ARD CATHETER PRODUCTS
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
jamie hartz
2400 bernville road
reading, PA 19605
MDR Report Key5701198
MDR Text Key46662977
Report Number1036844-2016-00306
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K900263
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2017
Device Catalogue NumberCS-25703-E
Device Lot Number23F15G0695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/02/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-