The maquet service territory manager (stm) evaluated the device and determined that the central handle fell off because the rivets fixing it to the light head were loosened.The most probable cause for this event is excessive mechanical effort that stressed the handle support ring during use.The volista user manual indicates to check the light head for any damage, on a daily basis prior to use.The replacement part is available, and the maquet stm will repair the defective equipment as soon as the customer allows access to the facility.
|