• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO ACTIVA; IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO ACTIVA; IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS) Back to Search Results
Model Number 37612
Device Problems Electromagnetic Interference (1194); Failure to Deliver Energy (1211); Device Displays Incorrect Message (2591)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 05/12/2016
Event Type  malfunction  
Event Description
Information was received from a patient who was implanted with a neurostimulator for dystonia.It was reported that a power on reset (por) message was seen following recharging; therapy was stopped.With assistance, the patient was able to clear the por.It was later reported that the patient saw another por the following day and he was back in the stone age, back before the dbs.The patient was again able to clear the por and turn therapy back on.It was also noted that there was a possible emi exposure as the patient recently had his teeth pulled.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key5705914
MDR Text Key46778772
Report Number3004209178-2016-11086
Device Sequence Number1
Product Code MRU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/14/2010
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/12/2016
Initial Date FDA Received06/08/2016
Supplement Dates Manufacturer Received05/12/2016
Supplement Dates FDA Received09/22/2017
Date Device Manufactured10/20/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
-
-