| Catalog Number |
H965750355 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
05/12/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It has been indicated by some hospital personnel that poor maintenance (e.G., too long between dressing changes) is responsible for the problems.Angiodynamics' investigation is on-going.Upon completion of the investigation, a supplemental medwatch will be submitted.((b)(4)) device discarded by hospital.
|
| |
|
Event or Problem Description
|
|
End user hospital is reporting a picc being removed & replaced due to central line-associated bloodstream infection (clabsi).The hospital has recently changed protocol of having picc team perform routine maintenance on piccs to having nurses perform the maintenance.The device was not retained.The current patient condition is reported as doing well.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history records was performed for the indicated packaging and picc assembly component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The march 2016 angiodynamics complaint report was reviewed for the bioflo picc product family and the failure mode "patient injury - infection." no adverse trend was identified.Labeling review: since the bioflo picc device was not returned for evaluation the root cause of the patient infection/allergic reaction and whether the picc was used in accordance with its labeling cannot be determined.The risk of the reported event failure mode - infection, is identified in the directions for use that is included with the bioflo picc device.The following is provided as a reference: "warning:contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged.If damage is found, call your angiodynamics representative.Inspect prior to use to verify that no damage has occurred during shipping." the root cause of the user hospital's reported complaint description cannot be confirmed as no sample was returned for evaluation.Review of the angiodynamics device history records showed all documents were in order.Potential contributing factors for the patient infections include the picc placement procedures and device care and maintenance.It was also reported that there was a change in personnel performing picc dressing changes that may have contributed to the patient infections.The sterilization records for the reported device lot were reviewed.The records were complete and in order.(b)(4).
|
| |
|
Search Alerts/Recalls
|