The surgeon indicated that the cause of the infection was unknown.There were no performance or any other type of adverse condition found with the pulse generator.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The pulse generator diagnostics were as expected for the programmed parameters.The battery, 2.944 volts as measured during testing shows an ifi=no condition and device memory revealed that 5.223% of the battery had been consumed.No obvious anomalies were noted with the returned lead portions and the condition of the returned lead portions was consistent with conditions that typically exist following an explant procedure.The majority of the lead assembly (body) including the electrode array section was not returned therefore an evaluation and resulting commentary cannot be made on that portion of the lead.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions revealed no discontinuities.
|