Taper ii.Medwatch sent to the fda on 06/13/2016.The reporter of the event was asked to inform apollo when explant surgery has been scheduled and to return the product for analysis.To date, apollo has not received an explant date or the device.Based on the serial number provided, the connecter type is assumed to be a taper ii.If returned, visual examination may confirm or determine another taper type associated with this event.Device labeling addresses the reported event of possible port/band leak as follows: precautions: care must be taken during band adjustment to avoid puncturing the tubing that connects the access port and band, as this will cause leakage and deflation of the inflatable section.Failure to create a stable, smooth path for the access port tubing, without sharp turns or bends, can result in tubing breaks and leakage.In order to avoid incorrect placement, the port should be placed lateral to the trocar opening.A pocket must be created for the port so that it is placed far enough from the trocar path to avoid abrupt kinking of the tubing.The tubing path should point in the direction of the access port connector so that the tubing will form a straight line with a gentle arching transition into the abdomen.Adverse events: unplanned deflation of the band may occur due to leakage from the band, the port or the connecting tubing.It is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.
|