• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SORIN HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SORIN HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-82
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/19/2016
Event Type  No Answer Provided  
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin heater-cooler system 3t.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the sorin heater-cooler system 3t system shut off during a case.The unit was replaced to complete the procedure.There was no report of patient injury.A sorin group field service representative was dispatched to the facility to investigate.The service representative found that the biomedical engineer had run the unit for 4 hours but was unable to reproduce the error.The service representative checked the power cord and plug and did not identify any issues.The connectors were all checked and reseated and the unit was tested for several hours.The issue was not duplicated.A technical safety inspection was performed and the unit was released to the customer.As the issue was not be reproduced, a root cause could not be determined and no corrective actions were identified.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.Evaluated on site by sorin service rep.
 
Event Description
Sorin group (b)(4) received a report that the sorin heater-cooler system 3t system shut off during a case.The unit was replaced to complete the procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key5721838
MDR Text Key47289393
Report Number9611109-2016-00361
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 05/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-82
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2016
Initial Date FDA Received06/13/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/22/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-