Certas plus valve was used for a lumbar peritoneal shunt implant.The patient at some point allegedly sat down, heard / felt a crack near the surgical site and reported having a headache approximately 20-30 minutes later.Upon explanting the certas plus valve, the surgeons found that it had cracked / tore in between the valve and siphonguard device.A new certas plus valve was implanted.
|
(b)(4).Upon completion of the investigation it was noted that the images were taken of the ¿as received¿ valve.The position of the cam when valve was received was at setting 3.The valve was hydrated for about 25 hours.The valve was visually inspected: and confirms the tear in the silicone housing.Review of the history device records confirmed the valve product code 82-8804lp, with lot cvcbfh, conformed to the specifications when released to stock in 10th march 2016.The root cause to the tear in the silicone housing could be due to the user, this however could not be determined.As noted in the ifu silicone has a low tear / cut resistance.Per hhe, it has been concluded that silicone housing tears are not design related, in order for the housing tear to occur the user has to compromise the silicone through a nick or tear in order for the event to occur.Validation testing demonstrates a robust design.Testing has shown that if the silicone housing has been compromised, a housing tear is likely to occur.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
|