Catalog Number 8065752101
Device Problem Break (1069)
Patient Problem No Code Available (3191)
Event Date 05/25/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that the ultrasound tip broke in the eye during an eye surgery.The tip was removed from the patient's eye and collected.The product was replaced and the procedure was completed without harm to the patient.A product sample has been requested for evaluation.
 
Manufacturer Narrative
A phaco tip sample was returned for evaluation.The front section of the tip was received still in the infusion sleeve.No lot number was identified with this complaint; therefore, a lot history review could not be conducted.A visual inspection of the phaco tip received was performed and deemed nonconforming.The tip is broken at the transition between the cone and cannula.The break area is slightly jagged with part of one section of the wall pushed inward.The wall thickness at the break area is even.Wear is present on the threads and back of flange.The complaint does confirm the phaco tip is broken.The exact reason of how and when the phaco tip became broken cannot be determined from this evaluation.The most likely reason for tip breakage is from a machine parameter not appropriate for this tip, tuning without water flow, occluded tip, or an improper sleeve installation.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
eddie darton, md, jd
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686660
MDR Report Key5727190
MDR Text Key48445578
Report Number1644019-2016-00924
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752101
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2016
Initial Date FDA Received06/15/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/30/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-