| Catalog Number |
NEXHD1552401 |
| Health Effect - Clinical Code |
Purulent Discharge (1812)
|
| Date of Event |
05/13/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was indicated that the device will not be returned for evaluation because the device has been disposed; therefore an analysis of the complaint device could not be completed.A review of the manufacturing documentation confirmed that the lot met the specification requirements.The review found no deviations or non-conformances that would have contributed to the reported complaint.Taking into consideration the evaluation conducted and the details of the complaint, this investigation was assigned the most probable root cause of an anticipated procedural complication.A complaint with a most probable root cause classification of an anticipated procedural complication indicates that a device related root cause does not apply and the complaint is due to a known effect of the procedure.
|
| |
|
Event or Problem Description
|
|
(b)(4): the nexsite device was successfully placed on (b)(6) 2015.The catheter was removed on (b)(6) 2015 as it was no longer required.The dacron was left in the subject and was reported to form a non-healing wound in the chest wall with purulent drainage.The dacron was removed on (b)(6) 2016 and the subject was given ceftazidime 200mg and vancomycin hci 1500mg at the dialysis unit.Blood cultures were drawn on (b)(6) 2016 and they were negative.There were no wound cultures taken.The patient had recovered by (b)(6) 2016.
|
| |
|
Search Alerts/Recalls
|