MAUDE Adverse Event Report: INVACARE FLORIDA OPERATIONS CS BED 9153650455; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE FLORIDA OPERATIONS CS BED 9153650455; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number 5410LOW
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available a supplemental record will be filed.
 
Event Description
The dealer stated that the weld on the head spring is broken.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the bed rails had been mounted in the wrong location causing the drive arm to bend, which confirmed the original complaint issue.However, as there was no malfunction of the device, this is no longer an fda reportable event.
 
Event Description
The dealer stated that the weld on the head spring is broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS BED 9153650455
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5728924
MDR Text Key47536085
Report Number1031452-2016-02642
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Replace
Type of Report Initial,Followup
Report Date 08/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number5410LOW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/09/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-