STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #6 11MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5531G611
Device Problems Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Loss of Range of Motion (2032); Injury (2348); Ambulation Difficulties (2544)
Event Date 05/20/2016
Event Type  Injury  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was retained by the patient and was not returned to the manufacturer.Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Patient had an inadequate range of motion.The doctor resected more distal femur, implanted a thinner tibial insert as well as changed the femur one size smaller.He used a posterior stabilized construction.This was the right side.There was nothing loose or broken.
 
Manufacturer Narrative
An event regarding poor range of motion involving a triathlon insert component was reported.The event was confirmed.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: medical review concluded: femoral component malposition with mismatch in flexion-extension gap space causing stiffness and requiring revision with exchange of femoral component and tibial bearing.Device history review: the device was manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar reported events for the lot referenced.Conclusion: clinician review of the medical records provided indicated that femoral component malposition with mismatch in flexion-extension gap space causing stiffness and requiring revision with exchange of femoral component and tibial bearing.No further investigation is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Patient had an inadequate range of motion.The doctor resected more distal femur, implanted a thinner tibial insert as well as changed the femur one size smaller.He used a posterior stabilized construction.This was the right side.There was nothing loose or broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INSERT #6 11MM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5732500
MDR Text Key47625007
Report Number0002249697-2016-01974
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2020
Device Catalogue Number5531G611
Device Lot NumberLEJ415
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/29/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
Patient Weight83
-
-