SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-01-00
Device Problems Temperature Problem (3022); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/23/2016
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Sorin group (b)(4) manufactures the sorin centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the pump of sorin centrifugal pump system with tubing clamp became warm and produced noise during a procedure.There was no report of patient injury.A sorin group field service representative was dispatched to the facility to investigate.The service representative confirmed the reported issue and identified a faulty gasket connector.The part was replaced and functional testing identified no further issues.The unit was released to the customer and the replaced part was scrapped.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.Discarded by 3rd-party service engineer.
 
Event Description
Sorin group (b)(4) received a report that the pump of sorin centrifugal pump system with tubing clamp became warm and produced noise during a procedure.There was no report of patient innjury.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The involved device was returned to livanova (b)(4) for further investigation and repair.The device was cleaned, disinfected and the insert for the clamp was replaced.A technical safety inspection was successfully completed before the device was returned to the customer.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr.25
munich,
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5740179
MDR Text Key47894453
Report Number9611109-2016-00415
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-