IRIS INTERNATIONAL IQ 200 SYSTEM; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Catalog Number 700-3325
Device Problem Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/08/2016
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) and customer technical specialist (cts) were at the customer site on (b)(4) 2016 and removed the barcode reader and installed another barcode reader.The repairs were verified per established service procedures.(b)(4).Related events: 2023446-2016-00284, bec internal identifier (b)(4).2023446-2016-00286, bec internal identifier (b)(4).
 
Event Description
The customer called to report dilutions were not being run correctly on their iq200 sprint urine microscopy analyzer.Erroneous patient results were not reported by the customer and there was no change or effect to patient treatment in connection to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ 200 SYSTEM
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer (Section G)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
gopal mohanty
9172 eton avenue
chatsworth, ca, CA 91311
8185277379
MDR Report Key5742924
MDR Text Key48000426
Report Number2023446-2016-00285
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-3325
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/08/2016
Initial Date FDA Received06/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-