| Model Number |
CIE2-15-24-00 |
| Medical Device Problem Codes |
Off-Label Use (1494); Improper or Incorrect Procedure or Method (2017); Malposition of Device (2616); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Hematoma (1884)
|
| Date of Event |
05/25/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Per the surgeon in the case, there was no device malfunction.Patient had very soft bone and should not have been implanted with the varilift-l device as a stand-alone.Supplemental fixation was added and the patient is doing fine post-revision surgery.The results of the investigation also note that the surgeon used 2 different size varilift-l implant devices at each level which is in contradiction to clearly stated requirements in the varilift-l surgical technique "bilateral devices implanted in the disc space must be of equal size.".
|
| |
|
Event or Problem Description
|
|
Revision surgery to add supplemental fixation for patient 3 weeks post varilift-l 2-level bilateral plif surgery due to wound hematoma and vertebral fracture.No device malfunction reported.
|
| |
|