PENUMBRA, INC. PENUMBRA SYSTEM ACE 64 REPERFUSION CATHETER; NRY Back to Search Results
Catalog Number 5MAXACE064-A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/26/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the penumbra system ace 64 reperfusion catheter (ace 64) was fractured approximately 1.0 cm from the hub.Conclusions: evaluation of the returned device revealed the ace 64 was fractured.This type of damage typically occurs due to improper handling during preparation.If the ace 64 is manipulated forcefully at extreme angles during removal from the packaging hoop or preparation for use, damage such as this may occur.Ace 64 devices are 100% visually inspected during in-process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a thrombectomy procedure, the hospital staff noticed that the penumbra system ace 64 reperfusion catheter (ace 64) was leaking saline at the junction of the hub and the catheter.There was no observed damage to the ace 64.However, due to the observed leak, the ace 64 was not used for the procedure.The procedure was completed using a new ace 64.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ACE 64 REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key5749089
Report Number3005168196-2016-00876
Device Sequence Number947251
Product Code NRY
UDI-Device Identifier00814548016207
UDI-Public00814548016207
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K152541
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/10/2018
Device Catalogue Number5MAXACE064-A
Device Lot NumberF64356
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2016
Initial Report FDA Received Date06/24/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-