MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS EPD 60000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS EPD 60000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.010
Device Problems Bent (1059); Moisture Damage (1405); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was duplicated and confirmed.The assignable root cause was determined to be due to improper handling, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported in the service order from (b)(6) that the electric pen drive device stopped working half way through an unspecified surgical procedure.During service and repair evaluation, it was observed that the device had bent coupling power supply pins were bent, made abnormal noise and had water stains on the motor surface.It was further observed that the device failed the following pre-tests: test for cable coupling, locking mechanism from cable coupling.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The initial medwatch listed the root cause as improper handling.It has been updated as improper maintenance, which is also user error/misuse/abuse.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPD 60000 RPM
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5750554
MDR Text Key49029771
Report Number8030965-2016-14057
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/26/2016
Initial Date FDA Received06/27/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-