(b)(4) was contacted about the cart and dispatched a service technician to be at the site the following day.On june 10, 2016, the technician arrived at the site and confirmed that the cart had overflowed with fluid.He then discovered upon further inspection that there was a large biomass inside of a cylinder.The technician removed the biomass and then verified that the cart was functioning as intended.The cart was then returned to service without further incident.
|
Ultra duo flex fluid cart, serial number (b)(4), was manufactured on october 6, 2014 and was 20 months old at the time this complaint was generated.The device history record (dhr) review noted no related non-conformances, requests for deviation (rfd), change notices (cn), or any other issues with manufacturing.The dhr review found no issues with the device and all verifications, inspections, and tests were successfully completed.Using crm to query for serial number (b)(4), the device was noted to have been previously repaired by zimmer biomet surgical once for a non-responsive monitor on (b)(6) 2016.The monitor was not associated with the reported event, so this was most likely a non-related issue.The reported event was confirmed by the service technician that performed the repair.On (b)(6) 2016, it was reported from (b)(6) that a cart overflowed during a case.On (b)(6) 2016, additional clarification from the customer revealed that the reported event occurred during surgery and that there was no patient involvement.There was no harm to the patient or operator associated with the reported event, no surgical procedure was affected, and the hospital staff was noted to be wearing personal protective equipment (ppe) at the time of the reported event.The customer also stated that the cart was not reading the correct fluid level due to a bolt clot holding down the level sensor.(b)(4) was contacted about the cart and dispatched a service technician to be at the site the following day.On april 10, 2016, the technician arrived at the site and confirmed that the cart had overflowed with fluid.He then discovered upon further inspection that there was a large biomass inside of a cylinder.The technician removed the biomass and then verified that the cart was functioning as intended.The cart was then returned to service without further incident.No repair checklist was required per customer relationship management (crm).Service work order (b)(4) on june 9, 2016.The root cause of the cart overflowing during a case was due to a large biomass coagulating inside of the cart.The instructions for use state after the cart is used, the cart needs to be transported to the evac so that it can be emptied and cleaned.If the customer would have foregone processing the cart and allowed fluid to sit inside of the cylinders, the biomaterial inside of the cylinders could begin to coagulate and congeal.The biomaterial could then prevent the fluid from emptying from the cylinder as well as prevent the level sensor float from measuring the fluid level inside of the cart.If biomaterial would have held the level sensor float down, when the customer would then attempt to operate the cart, the cart would not be able to properly shut off the cart when it would reach the 16.5 liter limit.The customer could then allow the cylinders to then overfill with fluid and therefore either spill out from the manifold housing or leak out of the cylinder through the vent line.The leaking fluid would then prevent the cart from operating as intended.Recommended actions: no recommended actions at this time.
|