• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS CAD II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS CAD II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 511.701
Device Problems Partial Blockage (1065); Power Conditioning Problem (1474); Device Operates Differently Than Expected (2913); Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Reporter's phone number: (b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that the compact air drive device trigger was jammed.During service and evaluation, it was observed that the device motor power was too low, trigger was blocked, and handle blocked.It was not reported if the event occurred during a surgical procedure, or if a spare device was available for use.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAD II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5752277
MDR Text Key49033293
Report Number8030965-2016-14067
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 06/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-